Element |
Wavelength (nm) |
Pneumatic Nebulizer |
U5000AT+ |
| Ag | 328.068 | 1 | 0.03 |
| Al | 396.152 | 2 | 0.06 |
| As | 189.042 | 3 | 0.7 |
| Ba | 233.527 | 0.5 | 0.01 |
| Be | 313.107 | 0.1 | 0.009 |
| Bi | 223.061 | 2 | 0.2 |
| Ca | 317.933 | 2 | 0.03 |
| Cd | 228.802 | 0.1 | 0.02 |
| Co | 228.616 | 0.2 | 0.02 |
| Cr | 267.716 | 0.2 | 0.01 |
| Cu | 324.752 | 0.6 | 0.02 |
| Fe | 238.204 | 0.1 | 0.02 |
| Mg | 285.213 | 0.5 | 0.06 |
| Mn | 257.610 | 0.1 | 0.03 |
| Mo | 202.030 | 0.6 | 0.3 |
| Ni | 231.604 | 0.4 | 0.06 |
| Pb | 220.353 | 2 | 0.2 |
| Sb | 206.833 | 2 | 0.3 |
| Se | 196.020 | 3 | 0.5 |
| Sn | 189.989 | 1 | 0.4 |
| Ti | 334.941 | 0.2 | 0.006 |
| Tl | 190.856 | 2 | 0.5 |
| V | 290.882 | 2 | 0.02 |
| Zn | 213.856 | 0.2 | 0.03 |
Notes: Detection limits are based on 3 sigma, with a 20-second integration time. Instrument used was a PerkinElmer 4300DV ICP-OES. |
| [close window] |